FDA Recall Terminated

GE Millennium Systems: MG. Multi-Geometry Digital CSE Dual Detector Gamma Camera. A modular nuclear medicine imaging system. The MG gantry allows the detectors to be oriented in the 180-degree and 101.25-degree orientations. The other modular core components include a 3-axis universal imaging table, a GENIE Acquisition station, dual Digital CSE (Correlated Signal Enhancement) detectors and high-performance collimators. Features optional real-time body-contoured SPECT and whole body scans, as well as ACuscan, an optional transmission attenuation correction system for SPeC and Gated SPECT scans.

Recall: Z-0414-06 · Initiated April 18, 2005

Recall

Recall Number
Z-0414-06
Event Number
33062
Firm
GE Healthcare
FEI Number
2126677
Product Code
KPS
Status
Terminated
Root Cause
Other
Initiated
April 18, 2005
Posted
January 24, 2006
Terminated
February 24, 2008
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Millennium Systems: MG. Multi-Geometry Digital CSE Dual Detector Gamma Camera. A modular nuclear medicine imaging system. The MG gantry allows the detectors to be oriented in the 180-degree and 101.25-degree orientations. The other modular core components include a 3-axis universal imaging table, a GENIE Acquisition station, dual Digital CSE (Correlated Signal Enhancement) detectors and high-performance collimators. Features optional real-time body-contoured SPECT and whole body scans, as well as ACuscan, an optional transmission attenuation correction system for SPeC and Gated SPECT scans.

Reason

In certain cases, one of the four collimator locking mechanisms that attach the collimator to the head may work itself free if collimator exchange has not been performed for an extended period of time. This situation, if neglected, could eventually cause the locking mechanism to unlock itself.

Action

Customer letter is being sent to all customers who have an affected system. The letter alerts customers of the potential release of the locking mechanism and requires a Weekly Collimator Exchange Procedure to avoid release of the locking latch until the permanent solution is made available.

Distribution

nationwide including Puerto Rico, and to countries including: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Denmark, Egypt, France, Germany, Iceland, India, Israel, Italy, Jordan, Korea, Kuwait, Luxembourg, Martinique, Mexico, New Zealand, Philippines Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Taiwan and United Kingdom.