FDA Recall Terminated

KCI BariAir Therapy System; AC-Powered Adjustable Hospital Bed; Bariatric Bed, manufactured by Kinetic Concepts, Inc., San Antonio, Texas

Recall: Z-0414-03 · Initiated October 21, 1998

Recall

Recall Number
Z-0414-03
Event Number
25197
Firm
Kinetic Concepts, Inc
FEI Number
1625774
Product Code
FNL
Status
Terminated
Root Cause
Other
Initiated
October 21, 1998
Posted
January 9, 2003
Terminated
January 28, 2003
Address
4958 Stout Drive, San Antonio, TX, 78219

Description

KCI BariAir Therapy System; AC-Powered Adjustable Hospital Bed; Bariatric Bed, manufactured by Kinetic Concepts, Inc., San Antonio, Texas

Reason

Electrical failure of the bed due to fluid egress into the junction box.

Action

In an October 21, 1998 e-mail the firm requested return of the product for a planned upgrade. The junction box cover was corrected at the same time of the upgrade.

Distribution

Nationwide

Quantity