TruWave Pressure monitoring set, REF: PX212: cc/12 in (30cm), PX260: (3 cc/60in (150cm), PX272: 3 cc/72in (180cm), PX284: 3 cc/84 in (210 cm), PX2X2: x2 (3cc), PX3X3284C: X3 (3cc), PX4X4: x4 (3cc), PX600F (3cc), PXMK1876: x 3 (3 cc), PXMK1940: (3 cc), PXMK2024: (3cc)/77 in (196cm),PXMK2041: (3 cc), PXMK2321: (3cc)/48in (122cm), PXMK2331: x4 (3cc)/12 in (30cm), PXMK9146: x 2 (3cc), PXMK1876: x3(3cc), PXVK1099:(3cc)/62in (157cm), PXMK0102: x2 (3cc)/84in (213cm), PXMK2043:X3(3CC), T001657A: x2 (3cc)/210cm, T001658A: x2 (3cc)/180 cm, T001660A: X3 (3CC)/150CM, T001709A: X2 (3 cc)/210 cm,T001767A: (3cc)/60in (150cm), T321573A: (3cc)/66 in (165cm), T433803A (3cc)/200cm, T441T01C: (3cc)/225 cm, T494C00B: (3 cc)/180 cm,
Recall
- Recall Number
- Z-0413-2024
- Event Number
- 93279
- Firm
- Edwards Lifesciences, LLC
- FEI Number
- 2015691
- Product Code
- DRS
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- October 4, 2023
- Posted
- November 29, 2023
- Address
- 1 Edwards Way, Irvine, CA, 92614-5688
Description
TruWave Pressure monitoring set, REF: PX212: cc/12 in (30cm), PX260: (3 cc/60in (150cm), PX272: 3 cc/72in (180cm), PX284: 3 cc/84 in (210 cm), PX2X2: x2 (3cc), PX3X3284C: X3 (3cc), PX4X4: x4 (3cc), PX600F (3cc), PXMK1876: x 3 (3 cc), PXMK1940: (3 cc), PXMK2024: (3cc)/77 in (196cm),PXMK2041: (3 cc), PXMK2321: (3cc)/48in (122cm), PXMK2331: x4 (3cc)/12 in (30cm), PXMK9146: x 2 (3cc), PXMK1876: x3(3cc), PXVK1099:(3cc)/62in (157cm), PXMK0102: x2 (3cc)/84in (213cm), PXMK2043:X3(3CC), T001657A: x2 (3cc)/210cm, T001658A: x2 (3cc)/180 cm, T001660A: X3 (3CC)/150CM, T001709A: X2 (3 cc)/210 cm,T001767A: (3cc)/60in (150cm), T321573A: (3cc)/66 in (165cm), T433803A (3cc)/200cm, T441T01C: (3cc)/225 cm, T494C00B: (3 cc)/180 cm,
Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer.
On 10/18/2023, Edwards Lifesciences issued a "Urgent: Medical device Correction Action via FedEx mail. Edwards ask consignees to take the following actions: 1. Follow the IFU for proper setup and flushing of your device. 2. Follow the instructions included in the enclosed customer acknowledgement form to complete the acknowledgement process. 3. Verify your inventory on the attached customer acknowledgement form. 3a.E-mail the completed form to Edwards Customer Service at [email protected], within 15 days from receipt of this notification. 4. Distribute this notice within your organization or to any organization where the potentially impacted product has been transferred. 5. Please post a copy of this notice near and/or with affected devices. 6. Contact Edwards Customer Service at 1-888-352-0904 if you would like to return any of your devices or have further questions. 7. Action Requested for Distributors: In addition to the actions above, please forward this customer communication to any of your customers who have purchased the impacted product.
US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA WA WI WV WY OUS: Argentina Australia Austria Bahrain Belgium Brazil Canada CANARY ISLANDS Chile Colombia Costa Rica Czech Republic Denmark Ecuador Faroe Islands France Germany Greece Guatemala Hong Kong Indonesia Ireland Israel Italy Japan Malaysia Netherlands New Zealand Northern Ireland Norway Panama Poland Portugal Saudi Arabia Singapore Slovakia Slovenia South Korea Spain Sweden Switzerland United Arab Emirates (UAE) United Kingdom Vietnam
118,570 units