FDA Recall Terminated

GE Millennium Systems: MC, MyoSight. Dual Head Gamma Camera. A compact cardiac-optimized nuclear imaging system. It is comprised of 2 digital CSE (Correlated Signal Enhancement) detectors in variable geometries mounted on a pass-through ring gantry, a compact imaging table, and a GENIE Acquisition station.

Recall: Z-0413-06 · Initiated April 18, 2005

Recall

Recall Number
Z-0413-06
Event Number
33062
Firm
GE Healthcare
FEI Number
2126677
Product Code
KPS
Status
Terminated
Root Cause
Other
Initiated
April 18, 2005
Posted
January 24, 2006
Terminated
February 24, 2008
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Millennium Systems: MC, MyoSight. Dual Head Gamma Camera. A compact cardiac-optimized nuclear imaging system. It is comprised of 2 digital CSE (Correlated Signal Enhancement) detectors in variable geometries mounted on a pass-through ring gantry, a compact imaging table, and a GENIE Acquisition station.

Reason

In certain cases, one of the four collimator locking mechanisms that attach the collimator to the head may work itself free if collimator exchange has not been performed for an extended period of time. This situation, if neglected, could eventually cause the locking mechanism to unlock itself.

Action

Customer letter is being sent to all customers who have an affected system. The letter alerts customers of the potential release of the locking mechanism and requires a Weekly Collimator Exchange Procedure to avoid release of the locking latch until the permanent solution is made available.

Distribution

nationwide including Puerto Rico, and to countries including: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Denmark, Egypt, France, Germany, Iceland, India, Israel, Italy, Jordan, Korea, Kuwait, Luxembourg, Martinique, Mexico, New Zealand, Philippines Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Taiwan and United Kingdom.

Quantity

1421 total for all affected Millenium Systems