FDA Recall Terminated

Smith & Nephew ACUFEX Rap-Pac ACL Accessory Kit, Sterile Catalog Number: 014725

Recall: Z-0413-03 · Initiated November 8, 2002

Recall

Recall Number
Z-0413-03
Event Number
25082
Firm
Smith & Nephew, Inc.
FEI Number
1219602
Product Code
HRX
Status
Terminated
Root Cause
Process control
Initiated
November 8, 2002
Posted
January 9, 2003
Terminated
April 20, 2005
Address
130 Forbes Boulevard, Mansfield, MA, 02048

Description

Smith & Nephew ACUFEX Rap-Pac ACL Accessory Kit, Sterile Catalog Number: 014725

Reason

Potential for non sterility as package tray does not contain aeration holes for ETO processing.

Action

Smith & Nephew notified customers by letter on 11/8/02. Users are requested to remove stock and return to Smith & Nephew.

Distribution

Nationwide: AL, CA, IL, KY, FL, GA, IN, ME, MI, NH,NY, OH, SC, TX, WV,

Quantity

77 units