FDA Recall Terminated

The Peripheral Cutting Balloon Device small Monorail Delivery System features a balloon with 3 or 4 atherotomes (microsurgical blades) mounted longitudinally on its outer surface. When the Peripheral Cutting Balloon device is inflated, the atherotomes score the plaque, creating initiation sites for crack propagation. Percutaneous Transluminal Angioplasty (PTA) with the Peripheral Cutting Balloon Device allows dilatation of the target lesion with less pressure.

Recall: Z-0410-06 · Initiated December 7, 2005

Recall

Recall Number
Z-0410-06
Event Number
34504
Firm
Boston Scientific Corporation
FEI Number
2024805
Product Code
LIT
Status
Terminated
Root Cause
Other
Initiated
December 7, 2005
Posted
February 1, 2006
Terminated
August 10, 2007
Address
3574 Ruffin Rd, San Diego, CA, 92123-2597

Description

The Peripheral Cutting Balloon Device small Monorail Delivery System features a balloon with 3 or 4 atherotomes (microsurgical blades) mounted longitudinally on its outer surface. When the Peripheral Cutting Balloon device is inflated, the atherotomes score the plaque, creating initiation sites for crack propagation. Percutaneous Transluminal Angioplasty (PTA) with the Peripheral Cutting Balloon Device allows dilatation of the target lesion with less pressure.

Reason

The distal shaft of the catheter may separate during withdrawal of the device from the patient. If the shaft separates, the surgical procedure may be prolonged or require more in-depth surgery to physically remove the broken-off shaft from the blood vessel.

Action

All affected direct accounts will be given notification by Overnight mail. A 100% of the total number of customer are to be contacted.

Distribution

Nationwide plus Argentina Austria Belgium Brunei Canada Chile China Colombia Czech Republic Denmark Finland France Germany Great Britain Hong Kong Hungary India Indonesia Ireland Israel Italy Latvia Lebanon Luxembourg Macau Malaysia Mexico Netherlands Norway Poland Portugal Puerto Rico Saudi Arabia. Singapore South Africa Spain Sweden Switzerland & Turkey

Quantity

individual qty in US:376, Worldwide distribution pending.