Software versions syngo CT VB20 running on the following Siemens SOMATOM CT Scanner. SOMATOM Force (Model #10742326), SOMATOM Definition As (Model #8098027), SOMATOM Definition Edge (Model #10590000), SOMATOM Definition Flash (Model #10430603), SOMATOM Drive (Model #10431700), SOMATOM Confidence (Model #10590100), and SOMATOM Edge Plus (Model # 10267000) with software syngo CT VB20
Recall
- Recall Number
- Z-0406-2020
- Event Number
- 84110
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- October 7, 2019
- Terminated
- November 14, 2022
- Address
- 40 Liberty Blvd, Malvern, PA, 19355-1418
Description
Software versions syngo CT VB20 running on the following Siemens SOMATOM CT Scanner. SOMATOM Force (Model #10742326), SOMATOM Definition As (Model #8098027), SOMATOM Definition Edge (Model #10590000), SOMATOM Definition Flash (Model #10430603), SOMATOM Drive (Model #10431700), SOMATOM Confidence (Model #10590100), and SOMATOM Edge Plus (Model # 10267000) with software syngo CT VB20
Software issue identified in the software versions syngo CT VB20 running on the SOMATOM CT Scanner: scan aborts and system crashes
Siemens Medical Solutions USA, Inc. issued a Customer Safety Advisory Notice (CSAN) to all affected customers via CT070/19/S. on 10/719 via email/or letter. The CSAN provides recommendations to avoid the issues until the technical solution can be applied. The issues will be resolved with software update version syngo CT VB20_SP1 (Service Pack 1). You will receive the updated software free of charge. Questions or require technical support, please contact the service organization at 1-800-888-7436.
Nationwide
305 units