Millar Disposable Angiographic Catheter (Left Heart), manufactured for Millar Instruments, Inc 6001-A Gulf Freeway, Houston, Texas 77023

Recall: Z-0401-2011 · Initiated October 14, 2010

Recall

Recall Number
Z-0401-2011
Event Number
57015
Firm
Millar Instruments, Inc
FEI Number
1625382
Product Code
DQO
Status
Terminated
Root Cause
Packaging
Initiated
October 14, 2010
Posted
November 17, 2010
Terminated
February 15, 2011
Address
6001-A Gulf Fwy, Houston, TX, 77023

Description

Millar Disposable Angiographic Catheter (Left Heart), manufactured for Millar Instruments, Inc 6001-A Gulf Freeway, Houston, Texas 77023

Reason

The sterile packaging was found to be compromised when tested after shipping conditions. This may result in infection or transmission of disease. A test for transportation of sterile package was conducted and resulted in several samples failing.

Action

Millar issued a Recall Report letter on October 14, 2010 to consignees identifying the affected product, sterility issue, and actions to be taken. Customers are to return all unused catheters to Millar for replacement. Customers can contact Millar about this recall at 832-667-7131.

Distribution

Worldwide Distribution: USA, including the states of MA, OH, PA, and D.C., and the countries of Canada and Germany.

Quantity

12 Units