FDA Recall Terminated

SKYLight Gamma Camera Gearbox brake, a component of the SKYLight Gamma Cameras 3/8" SPECT, Model number 4535 600 66661; and 5/8" SPECT, Model number 4535 602 20381, manufactured by Philips Medical Systems (Cleveland), Cleveland OH. Intended to produce images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.

Recall: Z-0398-2011 · Initiated October 15, 2010

Recall

Recall Number
Z-0398-2011
Event Number
57029
Firm
Philips Medical Systems
FEI Number
1525965
Product Code
KPS
Status
Terminated
Root Cause
Component change control
Initiated
October 15, 2010
Posted
November 17, 2010
Terminated
March 21, 2012
Address
3860 N 1st St, San Jose, CA, 95134-1702

Description

SKYLight Gamma Camera Gearbox brake, a component of the SKYLight Gamma Cameras 3/8" SPECT, Model number 4535 600 66661; and 5/8" SPECT, Model number 4535 602 20381, manufactured by Philips Medical Systems (Cleveland), Cleveland OH. Intended to produce images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.

Reason

Detector arm assembly may slide to hardware limit, potentially causing impact.

Action

Firm will notify consignees via Urgent - Medical Device Correction, dated Nov 3, 2010, via FedEx. The letter identified the affected product, described the problem, the hazard involved, how to identify affected products, actions to be taken, and actions planned by Philips. Customers are to contact their Field Safety Engineer immediately to schedule an inspection of their system. If customers suspect that the identified problem has occurred with their device, they should immediately stop using the system until Philips has confirmed it as safe to use. If customers need further information concerning this issue, they are to contact their local Philips representative.

Distribution

Worldwide Distribution -- USA and the Netherlands.

Quantity

24 units