Synthes Titanium Minimally Invasive Reduction Screw System Locking Cap, Part Number 04.631.502 Non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation.
Recall
- Recall Number
- Z-0396-2014
- Event Number
- 63230
- Firm
- Synthes USA HQ, Inc.
- FEI Number
- 3005180112
- Product Code
- NKB
- Status
- Terminated
- Root Cause
- Use error
- Initiated
- September 7, 2012
- Posted
- November 26, 2013
- Terminated
- December 18, 2015
- Address
- 1302 Wrights Ln E, West Chester, PA, 19380-3417
Description
Synthes Titanium Minimally Invasive Reduction Screw System Locking Cap, Part Number 04.631.502 Non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation.
This action is being initiated following a detailed investigation in response to reported complaints of post-operative locking cap loosening. Synthes has concluded that if the construct is mis-aligned during final tightening, the potential exists for locking caps to loosen post-operatively in certain situations.
Synthes sent an Urgent: Medical Device Product Removal letter on September 7, 2012, to all affected consignees. Consignees were requested to immediately cease using the product, examine their inventory and remove the parts from inventory. If they had the parts, they were further instructed to call Synthes at 1-800-479-6329 to obtain a Return Authorization Number, complete the verification section at the end of the letter and return the verification form. If they did not have the parts, they were requested to complete the verification section at the end of the letter and return the verification form. Consignees with questions were instructed to call 610-719-5450. For questions regarding this recall call 610-719-5000.
Nationwide Distribution including CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, MI, MN, MT, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI.
745