FDA Recall Terminated

Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments labeled as System 1000, AltraTouch 1000, Baxter Tina and Baxter Aurora; Baxter Healthcare Corporation, Deerfield, IL 60015 USA or Baxter Healthcare Corporation, McGaw Park, IL 60085 USA

Recall: Z-0395-05 · Initiated May 28, 2003

Recall

Recall Number
Z-0395-05
Event Number
30486
Firm
Baxter Healthcare Renal Div
FEI Number
3000210122
Product Code
FKP
Status
Terminated
Root Cause
Other
Initiated
May 28, 2003
Posted
December 2, 2004
Terminated
February 26, 2008
Address
1620 Waukegan Rd, Bldg R, Mc Gaw Park, IL, 60085-6730

Description

Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments labeled as System 1000, AltraTouch 1000, Baxter Tina and Baxter Aurora; Baxter Healthcare Corporation, Deerfield, IL 60015 USA or Baxter Healthcare Corporation, McGaw Park, IL 60085 USA

Reason

The air detector may not detect air bubbles consistently at the selected limit.

Action

Important Product Information letters were sent to customers on 5/28/03, informing them of the potential for the air detector not detecting air bubbles consistently. The letters instructed the accounts to measure the air detector sensor voltage output on each of their instruments, using the enclosed instructions, to determine whether the installation of an air detector attenuator connector is needed. The instructions also included directions for installation of the attenuator connector. Since the instrument requires re-qualification after the testing and installation, the accounts were given the choice to have the testing and installation done during the next service call or as part of the instrument''s next normally scheduled maintenance.

Distribution

The devices were distributed nationwide and internationally to Baxter subsidiaries in Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Denmark, Ecuador, Finland, France, the Republic of Georgia, Greece, Guatemala, Honduras, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Korea, Lebanon, Malaysia, Morocco, the Netherlands, Oman, Palestine, Panama, Peru, the Philippines, Poland, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Taiwan, Thailand, Tunisia, Turkey and the United Kingdom.

Quantity

28,482 units