Biomet Microfixation RapidFlap System, 12MM RapidFlap SpinDown Clamp, 12 Pack Ref Number: 75-1020-12 Used for the re-attachment of the bone flap after a craniotomy.
Recall
- Recall Number
- Z-0394-2019
- Event Number
- 81072
- Firm
- Zimmer Biomet, Inc.
- FEI Number
- 1825034
- Product Code
- GXN
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- August 22, 2018
- Terminated
- June 17, 2021
- Address
- 56 E Bell Dr, Warsaw, IN, 46582-6989
Description
Biomet Microfixation RapidFlap System, 12MM RapidFlap SpinDown Clamp, 12 Pack Ref Number: 75-1020-12 Used for the re-attachment of the bone flap after a craniotomy.
Spin Down RapidFlap nut that interfaces with the outer plate was assembled incorrectly.
Zimmer Biomet conducted recall to the consumer/user level, including any intermediate wholesale or retail consignees. Distributors notified initially via email on 8/22/18. The firm expanded the recall on 10/5/18 to include all lots with lot expiry before 09/18/2023, letters issued via email and FedEx to Distributors/ Risk Managers/Surgeons. Letter identifies issue and requesting removing and return of product. Surgeons will be provided with a letter that identifies the issue and their responsibilities, which include: - Following any specified patient monitoring instructions - Returning Certificate of Acknowledgment to Zimmer Biomet. Hospital risk managers will be provided with a letter identifying the issue. Questions or concerns after reviewing this notice, please call customer service at 907-741-4400 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected] .
Worldwide distribution: US Nationwide, Canada Argentina Australia Belgium Brazil Chile Colombia Costa Rica Czech Republic Dominican Republic Ecuador Finland France China Ecuador Germany Greece Hong Kong Indonesia Italy Japan Jordan Korea Lebanon Malaysia Netherland Paraguay Peru Singapore South Africa Spain Sweden Switzerland Thailand Trinidad and Tobago Turkey United Arab Emirates United Kingdom Uruguay Vietnam
11736 units