BD MAX" CT/GC/TV. Catalog Number(s): 443824,442970,443590 (RUO).for collection and transport of vaginal and endocervical swab specimens, in addition to the preservation of urine from males and females.
Recall
- Recall Number
- Z-0394-2018
- Event Number
- 78808
- Firm
- Becton Dickinson & Co.
- FEI Number
- 1119779
- Product Code
- NSU
- Status
- Terminated
- Root Cause
- Use error
- Initiated
- July 18, 2017
- Terminated
- July 9, 2018
- Address
- 7 Loveton Cir, Sparks, MD, 21152-9212
Description
BD MAX" CT/GC/TV. Catalog Number(s): 443824,442970,443590 (RUO).for collection and transport of vaginal and endocervical swab specimens, in addition to the preservation of urine from males and females.
BD recently notified customers in July 2017 regarding an increased rate of unresolved (UNR) and indeterminate (IND) results for the BD MAXTM Vaginal Panel, experienced by some customers. A non-reportable result may lead to repeat of the test and/or re-collection of a specimen.
Customers were notified via UPS on 07/18/2017. The recall for BD MAX" Vaginal Panel and BD MAX" UVE Specimen Collection Kit was initiated on July 18, 2017 and expanded to include BD MAX" CT/GC/TV on August 11, 2017. The recall expansion letter to customers was dated 08/14/2017.
US and OUS.