Truform Therapeutic Classic Medical Style Compression Pantyhose 20-30 mmHG, A Div of SAI, Cincinnati, OH 45209 Product Usage: To help relieve tired, heavy legs, and moderate to pronounced varicose veins and swelling.
Recall
- Recall Number
- Z-0392-2012
- Event Number
- 60371
- Firm
- Surgical Appliance Industries
- FEI Number
- 1511629
- Product Code
- FQL
- Status
- Terminated
- Root Cause
- Employee error
- Initiated
- October 7, 2011
- Posted
- January 11, 2012
- Terminated
- March 5, 2013
- Address
- 3960 Rosslyn Dr, Cincinnati, OH, 45209
Description
Truform Therapeutic Classic Medical Style Compression Pantyhose 20-30 mmHG, A Div of SAI, Cincinnati, OH 45209 Product Usage: To help relieve tired, heavy legs, and moderate to pronounced varicose veins and swelling.
During an audit with the FDA, it was discovered that a new compression testing device used for testing hosiery had been implemented and that the operators were not properly trained. Additionally, the acceptance criteria for the new device were not changed from the old testing device to reflect additional data (compression profile percentages). This resulted in the device history files containing s
The firm, Surgical Appliance Industries sent an URGENT: MEDICAL DEVICE RECALL letters dated November 7, 2011 to all affected customers. The letter described the products, problem and actions to be taken. Customers were instructed to immediately examine their inventory and quarantine all product(s) subject to recall for return and exchange. In addition, the letter states that if the product was further distributed to contact their customers and notify them of the recall. Customers were asked to complete and return the enclosed Recall Response Form. For any questions contact Quality Systems Manager, at 513-271-4594 ext. 282.
Worldwide Distribution - USA (nationwide) and the countries of: Australia, Canada, Colombia, Hong Kong, Peru, Singapore, Spain, and Venezuela.
11819 pairs