FDA Recall Terminated

Brilliance iCT Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Recall: Z-0388-2018 · Initiated November 10, 2017

Recall

Recall Number
Z-0388-2018
Event Number
78852
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAK
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
November 10, 2017
Terminated
November 5, 2020
Address
595 Miner Rd, Cleveland, OH, 44143-2131

Description

Brilliance iCT Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Reason

Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.

Action

An Urgent - Field Safety Notice Medical Device Correction letter, dated November 1, 2017, was sent to customers. The letter identified the affected device, described the issue, and discussed the hazard involved. Users are to cease use immediately and contact their Philips Field Service Engineer, if any unusual noises are heard within the gantry, or if the patient support tabletop is in free float. Philips Field Service Engineers will correct the system by replacing the affected fasteners. If further info or support concerning this issue is needed, customers can contact their local Philips rep or the Customer Care Solutions Center at 1-800-722-9377.

Distribution

Worldwide Distribution -- US, Argentina, China, Republic of Korea, Russian Federation, Slovakia, Turkey, and United Kingdom. 4 in US, (1 military), 14 foreign accounts

Quantity

13