Sabraset Administration Sets with AutoClamp; for use with model 6060 multi-therapy pumps only; a prescription, sterile, non-pyrogenic fluid pathway; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., the following sets were made in Mexico: a) Sabraset 56050 100 Administration Set with 100 mL Bag and Cassette, 56'', Volume 3 mL, 100 mL Bag with Fill Port, Cassette with AutoClamp Device, Slide Clamp, Male Luer Lock Adapter b) Sabraset 56050 250 Administration Set with 2500 mL Bag and Cassette, 53'', Volume 3 mL, 250 mL Bag with Fill Port, Cassette with AutoClamp Device, Slide Clamp, Male Luer Lock Adapter
Recall
- Recall Number
- Z-0386-05
- Event Number
- 30506
- Firm
- Baxter Healthcare Corp. Rt.
- FEI Number
- 1416980
- Product Code
- FPA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 19, 2004
- Posted
- January 8, 2005
- Terminated
- June 28, 2006
- Address
- 120 & Wilson Rd, Round Lake, IL, 60073
Description
Sabraset Administration Sets with AutoClamp; for use with model 6060 multi-therapy pumps only; a prescription, sterile, non-pyrogenic fluid pathway; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., the following sets were made in Mexico: a) Sabraset 56050 100 Administration Set with 100 mL Bag and Cassette, 56'', Volume 3 mL, 100 mL Bag with Fill Port, Cassette with AutoClamp Device, Slide Clamp, Male Luer Lock Adapter b) Sabraset 56050 250 Administration Set with 2500 mL Bag and Cassette, 53'', Volume 3 mL, 250 mL Bag with Fill Port, Cassette with AutoClamp Device, Slide Clamp, Male Luer Lock Adapter
A missing platen assembly in the cassette of the solution sets results in the infusion pump being unable to occlude the tubing of the set. This can lead to free-flow of the solution and a possible overdose of medication to the patient.
Urgent Product Recall letters dated 11/19/04 were sent to the direct accounts on the same date via UPS ground. The accounts were informed that free-flow situations were associated with the listed product codes and lot numbers due to a missing platen assembly in the cassette of the solution sets, which could result in an overdose of medication to the patient. The accounts were requested to stop use of the affected lot numbers and product codes and return the sets to Baxter c/o NNC Group, 2670 Executive Dr., Indianapolis, IN 46241, following the enclosed instructions. Any questions concerning the return process were directed to NCC at 1-866-300-5085. Any medical or technical questions were directed to Baxter at 1-800-422-9837.
Nationwide, including Puerto Rico, and internationally to Australia, Belgium, Canada, Hong Kong, Israel, Japan, Singapore, Sweden and Switzerland
1,355,820 sets