FDA Recall
Terminated
Clinical Chemistry Prealbumin, List Number 1E02-20
Recall: Z-0385-06
·
Initiated November 10, 2005
Recall
- Recall Number
- Z-0385-06
- Event Number
- 34281
- Firm
- Abbott Laboratories Diagnostic Div
- FEI Number
- 2018433
- Product Code
- JZJ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 10, 2005
- Posted
- January 11, 2006
- Terminated
- March 20, 2007
- Address
- 820 Mission St, South Pasadena, CA, 91030-3142
Description
Clinical Chemistry Prealbumin, List Number 1E02-20
Reason
These specific protein assays are not meeting the 57-day on board stability claim
Action
A Device Correction letter identifying the changes being made to the frequency of QC monitoring for these assays, will be distributed to all AEROSET and ARCHITECT c8000 customers which utilize the IgM, Prealbumin and Apolipoprotein Al Specifc Protein assays on November 10,2005 via direct mailing. Outside the U.S. each country organization will need to determine the appropriate method for the effectiveness check..
Distribution
Nationwide, plus Australia, Canada, Columbia, Germany, Guatemala, Hong Kong, mexico, & New Zealand
Quantity
734 (568 in US)