FDA Recall Terminated

Roche Elecsys total PSA CalSet, catalog number 11731696322.

Recall: Z-0382-06 · Initiated December 7, 2005

Recall

Recall Number
Z-0382-06
Event Number
34273
Firm
Roche Diagnostics Corp.
FEI Number
1823260
Product Code
LTJ
Status
Terminated
Root Cause
Other
Initiated
December 7, 2005
Posted
January 11, 2006
Terminated
February 26, 2006
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

Roche Elecsys total PSA CalSet, catalog number 11731696322.

Reason

There is the possibility of high results with sera expected to have non-detectable levels of tPSA if these reagent lots are used in combination on the Modular Analytics E170 analyzer.

Action

Consignees were notified of the incompatibility via reagent bulletin dated 12/7/05.

Distribution

Nationwide.

Quantity

25,000 kits