FDA Recall Terminated

Ventilator System Servo-i V3.1 Maquet Critical Care PC1771

Recall: Z-0379-2007 · Initiated May 23, 2006

Recall

Recall Number
Z-0379-2007
Event Number
37099
Firm
MAQUET Inc.
FEI Number
3008355164
Product Code
CBK
Status
Terminated
Root Cause
Other
Initiated
May 23, 2006
Posted
February 10, 2007
Terminated
March 23, 2009
Address
1140 US Highway 22, Bridgewater, NJ, 08807-2958

Description

Ventilator System Servo-i V3.1 Maquet Critical Care PC1771

Reason

The Connection of certain nebulizers to the built in driver has lead to reduced or no nebulization effect on the Servo-i Ventilator

Action

Maquet Critical Care issued an ''Extended Warranty'' letter on 5/23/2006. On March 13, 2007, Maquet modified its recall strategy and issued a Recall Notification letter accompanied by a Safety Alert letter to all customers with Servo-i ventilators, serial numbers in the 14000-22772 range, with Servo nebulizer installed. The Safety Alert letter provides customers with information on how to identify if the unit is affected, and how to identify if the unit has shown any failures, as well as, any precautions. Firm initiated recall is on-going

Distribution

Worldwide

Quantity

2,436 units