FDA Recall Open, Classified

SAFE-T-FILL Micro Capillary Blood Collection, 125 L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection, 300 L Prepared with Dipotassium EDTA Purple, Model: 07 7053; SAFE-T-FILL Micro Capillary Blood Collection, 200 L Prepared with Dipotassium EDTA Purple, Model: 07 7056; SAFE-T-FILL Micro Capillary Blood Collection, 200 L Prepared with Dipotassium EDTA Flat Bottom Microtube, Purple, Model: 07 7058 SAFE-T-FILL End-to-End Capillary Tubes, 30 L; Plastic, Prepared with EDTA Dipotassium Model: 06 0910

Recall: Z-0372-2026 · Initiated May 5, 2025

Recall

Recall Number
Z-0372-2026
Event Number
97551
Firm
ASP Global, LLC. dba Anatomy Supply Partners, LLC.
FEI Number
3004824601
Product Code
JKA
Status
Open, Classified
Root Cause
Component design/selection
Initiated
May 5, 2025
Posted
October 27, 2025
Address
7800 Third Flag Pkwy, Austell, GA, 30168-7657

Description

SAFE-T-FILL Micro Capillary Blood Collection, 125 L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection, 300 L Prepared with Dipotassium EDTA Purple, Model: 07 7053; SAFE-T-FILL Micro Capillary Blood Collection, 200 L Prepared with Dipotassium EDTA Purple, Model: 07 7056; SAFE-T-FILL Micro Capillary Blood Collection, 200 L Prepared with Dipotassium EDTA Flat Bottom Microtube, Purple, Model: 07 7058 SAFE-T-FILL End-to-End Capillary Tubes, 30 L; Plastic, Prepared with EDTA Dipotassium Model: 06 0910

Reason

All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.

Action

Starting on 5//5/2025 customers were informed that there had been reports of falsely elevated blood lead levels associated with the use of affected micro capillary tubes for lead testing when used with Magellan LeadCare Testing Systems. On 9/8/2025, recall notices were emailed to customers who were asked to do the following: Discontinue use of the recalled products immediately. b) Quarantine any remaining inventory to prevent further use. c) Dispose remaining inventory in accordance with local regulations or notify firm at [email protected] for return authorization and disposal assistance. d) Complete and return the enclosed Recall Response Form to confirm receipt and actions to [email protected] Distributors are asked to notify their customers of the recall or to forward their customer information to the recalling firm so they might notify these customers.

Distribution

Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI and the country of CA.

Quantity

4,272,858