FDA Recall
Terminated
ACCU-CHEK Inform meter; catalog No. 3035123.
Recall: Z-0372-05
·
Initiated November 19, 2004
Recall
- Recall Number
- Z-0372-05
- Event Number
- 30462
- Firm
- Roche Diagnostics Corp.
- FEI Number
- 1823260
- Product Code
- NBW
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 19, 2004
- Posted
- December 28, 2004
- Terminated
- June 7, 2005
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025
Description
ACCU-CHEK Inform meter; catalog No. 3035123.
Reason
A firmware bug may allow test results from other systems, recorded into the INFORM system via the OTE functionality, to be reversed from positive to negative, and vice versa.
Action
Consignees were notified via a letter dated 11/19/04, which instructed them to discontinue using the OTE functionality.
Distribution
Nationwide.
Quantity
4697