FDA Recall Terminated

CORFLO* Percutaneous Endoscopic Gastrostomy (PEG) Kit, Labeled as CORFLO* PEG Kit with ENFit Connector. 2 units per case The devices could be branded as Halyard Health.

Recall: Z-0371-2020 · Initiated October 8, 2019

Recall

Recall Number
Z-0371-2020
Event Number
84021
Firm
Avanos Medical, Inc.
FEI Number
3011270181
Product Code
KNT
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
October 8, 2019
Terminated
March 16, 2022
Address
5405 Windward Pkwy, Alpharetta, GA, 30004-3894

Description

CORFLO* Percutaneous Endoscopic Gastrostomy (PEG) Kit, Labeled as CORFLO* PEG Kit with ENFit Connector. 2 units per case The devices could be branded as Halyard Health.

Reason

The connector piece to the PEG tube does not pass over the guidewire.

Action

Avanos Medical notified customers of the recall on about 10/08/2019, via "URGENT: PRODUCT RECALL" letter. Customers were informed that there is a potential where the PEG tube does not pass over the guidewire. Instructions included to evaluate your inventory for any affected products, discontinue use of any affected product, quarantine any on hand, and arranged for the return of affected product. Customers were also instructed to complete the provide Recall Acknowledgement Form (Attachment 1) and return a copy of the form to Avanos in the USA by email to [email protected] or by FAX to 1-888-807-1084. Outside the USA, please send by email to [email protected] or by FAX to +44 (0) 20 7660 1462. For further assistance, please contact Avanos by email at [email protected] or by FAX to 1-888-807-1084. Outside the USA, please send by email to [email protected] or by FAX to +44 (0) 20 7660 1462. You may also contact Avanos Customer Service at [email protected].

Distribution

Distribution to CT, IN, MD, NJ, PA, VA. International distribution to Great Britain.

Quantity

104 cases (208 units) total