Medcare Products Max-500 Stand, Total Support, patient stand, model #400006. The responsible firm name listed in the Operations Manual is Medcare Products, Burnsville, MN. Used for assisting patients/residents to a standing position.
Recall
- Recall Number
- Z-0370-2014
- Event Number
- 66056
- Firm
- Ergosafe Products LLC DBA Prism Medical Services USA
- FEI Number
- 3007802293
- Product Code
- FSA
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- October 30, 2013
- Posted
- November 20, 2013
- Terminated
- May 17, 2017
- Address
- 45 Progress Pkwy, Maryland Heights, MO, 63043-3701
Description
Medcare Products Max-500 Stand, Total Support, patient stand, model #400006. The responsible firm name listed in the Operations Manual is Medcare Products, Burnsville, MN. Used for assisting patients/residents to a standing position.
The lift might fail with the labeled 500-lb. load rating.
The firm, Medcare Products, sent an "URGENT FIELD CORRECTIVE ACTIONS" letter dated October 21, 2013 via certified mail on October 30, 2013 to its customers. The letter described the product, problem and action to be taken. The customers were instructed to peel off the boom stickers and serial number label and replace with the new ones supplied; review the revised Manual and dispose of the original; and complete the action report form once the products have been relabeled. Please refer to your Operations Manual or visit our website at www.medcarelifts.com to view the manual online. If we can answer any questions that you may have regarding the testing standards or the relabeling of the products, please do not hesitate to contact us at 1-800-695-4479.
Nationwide distribution.
97 units