FDA Recall
Terminated
Alsius CoolGard Temperature Regulation System
Recall: Z-0369-06
·
Initiated November 9, 2005
Recall
- Recall Number
- Z-0369-06
- Event Number
- 34073
- Firm
- Alsius Corporation
- FEI Number
- 3002922896
- Product Code
- NCX
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 9, 2005
- Posted
- January 10, 2006
- Terminated
- February 5, 2007
- Address
- 15770 Laguna Canyon Rd, Ste 150, Irvine, CA, 92618-3111
Description
Alsius CoolGard Temperature Regulation System
Reason
Alsius has found that fluid spills or ingress may result in an alarm condition. The potential health risk involved in this state can be the discontinuance of therapy.
Action
Alsius Corporation is notifying customers of an issue requiring service of their CoolGard 3000 by letter shipped via FEDEX and DHL. Verification of receipt of the notification letter to level A will be undertaken by using tracked delivery requiring signature upon receipt.
Distribution
Nationwide and in Uruguay, Netherlands, Switzerland, Spain, Poland, China, Norway, Malaysia, Italy, Ireland, Greece, Germany, France, Finland, Canada, Belgium, Austria, Bahamas, and United Kingdom
Quantity
239