FDA Recall Terminated

Biosense Webster NaviStar RMT Diagnostic/Ablation Steerable Tip 4mm Catheters; Part Numbers: D-1257-01 & D-1257-02

Recall: Z-0368-2010 · Initiated October 2, 2008

Recall

Recall Number
Z-0368-2010
Event Number
50784
Firm
BioSense Webster, Inc. Circuito Interior Norte, # 1820 Parque Industrial Salvarcar Cd. Juarez, Chihuahua Mexico
FEI Number
3003248396
Product Code
LPB
Status
Terminated
Root Cause
Process control
Initiated
October 2, 2008
Posted
November 19, 2009
Terminated
November 23, 2009

Description

Biosense Webster NaviStar RMT Diagnostic/Ablation Steerable Tip 4mm Catheters; Part Numbers: D-1257-01 & D-1257-02

Reason

Biosense Webster initiated the recall because of the potential concern with one of the Printed Circuit Board (PCB) components used in the NaviStar RMT Catheters, which may cause the ablation catheter icon to shift on the map viewer when RF energy is being applied. Therefore the user, attempting a linear ablation in the location displayed on the CARTO RMT screen, may ultimately apply RF energy to u

Action

The firm sent an "Urgent Medical Device Correction" letter to consignees on October 6, 2008 informing consignees. Users were advised to pay "close attention to the location of the ablation catheter icon on the map viewer when applying RF energy." If consignees suspect that they have a defective NAVISTAR RMT Catheter, they were instructed to notify Biosense Webster and to return the defective catheter for evaluation and refund/replacement. Customer were instructed to contact their local Biosense Webster representative for any questions or further information. No return reply was requested.

Distribution

CA, CT, FL, GA, IA, IL, IN, KS, KY, MA, MI, MN, MO, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI

Quantity

2559 units