FDA Recall
Terminated
GE Healthcare OEC 9900 mobile fluoroscopic x-ray system
Recall: Z-0368-2008
·
Initiated December 14, 2007
Recall
- Recall Number
- Z-0368-2008
- Event Number
- 49536
- Firm
- OEC Medical Systems, Inc
- FEI Number
- 1720753
- Product Code
- JAA
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- December 14, 2007
- Posted
- September 29, 2008
- Terminated
- April 13, 2012
- Address
- 384 Wright Brothers Dr, Salt Lake City, UT, 84116-2862
Description
GE Healthcare OEC 9900 mobile fluoroscopic x-ray system
Reason
Possible overexposure: Failure to apply the entrance exposure rate (EER) tube current limit calibration to the automatic exposure rate (AER) control system when the anatomical profile mode is changed from the default selection to another selection. Manual technique modes are not affected.
Action
GE OEC sent a letter to customers December 15, 2007, and scheduled field representative visits for corrective software solution installation.
Distribution
Worldwide distribution: USA, Finland, Netherlands, Germany, Switzerland, France, United Kingdom, Australia, Italy, Ireland, Spain, Sweden, Belgium, New Zealand, China, Singapore, Poland, Norway, and India.
Quantity
710