FDA Recall Terminated

Zimmer ZMR Femoral Spout & Cone Bodies and Trabecular Metal Hip Prosthesis Femoral Stems: FEMORAL BODY REVISION NITRIDED POROUS 12/14 NECK TAPER EXTENDED 46 MM NECK OFFSET SPOUT BODY SIZE E 55 MM BUILD-UP TIVANIUM TI-6AL-4V ALLOY STERILE. Item number: 00-9991-021-55

Recall: Z-0367-2012 · Initiated November 16, 2011

Recall

Recall Number
Z-0367-2012
Event Number
60419
Firm
Zimmer Inc.
FEI Number
1000220733
Product Code
LPH
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
November 16, 2011
Terminated
August 28, 2013
Address
345 E Main St, Warsaw, IN, 46580-2746

Description

Zimmer ZMR Femoral Spout & Cone Bodies and Trabecular Metal Hip Prosthesis Femoral Stems: FEMORAL BODY REVISION NITRIDED POROUS 12/14 NECK TAPER EXTENDED 46 MM NECK OFFSET SPOUT BODY SIZE E 55 MM BUILD-UP TIVANIUM TI-6AL-4V ALLOY STERILE. Item number: 00-9991-021-55

Reason

Lack of assurance of proper metal fatigue strength due to metal grain structure anomaly. In 2006 a recall was initiated as a result of the grain anomaly; however, these 4 lots had been split off during normal production and had not been included in the 2006 recall.

Action

On 11/16/11, Zimmer sent letters titled "URGENT MEDICAL DEVICE RECALL" to distributors via electronic mail and to risk managers and surgeons via FedEx indicating that the product must be immediately removed from use and quarantined. " Stop using the affected lots and place affected product in quarantine. " Review available product on the shelf " Identify those part/lot combinations listed in the recall communication " Immediately remove any implants found that match part/lot combinations listed in the recall communication and quarantine " Return affected lots and signature of acknowledgement to Zimmer

Distribution

Worldwide distribution: USA (NY), AUSTRALIA, SWITZERLAND, and DENMARK

Quantity

1 unit (International Only Switzerland)