Philips Medical Systems, Cardiac Viewer or Pulmonary Viewer Application that can run on all Brilliance CT scanners using software versions 2.0 through V 2.2.5 and all EBW workstations using software versions 2.0 through V3.5.2. Philips Medical Systems, Cleveland, OH 44143. (The product is a software application that is installed in Philips Medical System CT scanners.)
Recall
- Recall Number
- Z-0367-2008
- Event Number
- 44790
- Firm
- Philips Medical Systems (Cleveland) Inc
- FEI Number
- 1525965
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- September 10, 2007
- Posted
- December 29, 2007
- Terminated
- September 19, 2012
- Address
- 595 Miner Road, Cleveland, OH, 44143-2131
Description
Philips Medical Systems, Cardiac Viewer or Pulmonary Viewer Application that can run on all Brilliance CT scanners using software versions 2.0 through V 2.2.5 and all EBW workstations using software versions 2.0 through V3.5.2. Philips Medical Systems, Cleveland, OH 44143. (The product is a software application that is installed in Philips Medical System CT scanners.)
Incorrect Measurement: Incorrect measurement will occur when the operator uses the distance or area measurement function, such as Graphic Tool line or ROI on zoomed and batch saved images from the Cardiac Viewer or Pulmonary Viewer and then views them in the CT Viewer or on a PACS System.
Consignees were notified of this recall by an Urgent Device Notification letter sent on 9/10/07. The letter instructed the users to not save any zoomed images to a batch file. Make all measurements of zoomed images in the CT viewer, which provides correct measurements. Do not perform measurements on any zoomed batched images that have been stored on a PACS System. Phillips in investigating the problem and will provide updates free of charge when completed.
Worldwide Distribution - USA including states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WI, WV, and WY, and country of Canada.
2,544