SoftLab GUI Version 4.0.1.0 - 4.0.1.16, 4.0.2.0 - 4.0.2.10, 4.0.3.0 - 4.0.3.13, and 4.0.4.0 - 4.0.4.5. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. Manufacture/Distribution Dates: 4.0.1.0 - 08/06/04, 4.0.1.4 - 10/18/04, 4.0.1.5 - 01/18/04, 4.0.1.6 - 12/15/04, 4.0.1.7 - 01/10/05, 4.0.1.8 - 03/29/05, 4.0.1.9 - 05/06/05, 4.0.1.13 - 01/25/06, 4.0.1.14 - 03/26/07 4.0.1.15 - 05/02/07, 4.0.1.16 - 08/22/08, 4.0.2.0 - 04/08/05, 4.0.2.1 - 08/13/08, 4.0.2.2 - 06/06/05, 4.0.2.3 - 09/19/05, 4.0.2.4 - 09/19/05, 4.0.2.5 - 09/19/05, 4.0.2.7 - 02/13/06, 4.0.2.8 - 12/12/06, 4.0.2.9 - 02/20/08, 04.2.10 - 08/13/08, 4.0.3.2 - 05/17/06, 4.0.3.3 - 09/18/06, 4.0.3.4 - 09/27/06, 4.0.3.5 - 11/29/06, 4.0.3.6 - 06/22/07, 4.0.3.7 - 03/28/07, 4.0.3.8 - 06/22/07, 4.0.3.9 - 08/03/07, 4.0.3.10 - 08/24/07, 4.0.3.11 - 09/24/07, 4.0.3.12 - 03/07/08, 4.0.3.13 - 06/13/08, 4.0.4.2 - 02/22/08, 4.0.4.3 - 05/22/08, 4.0.4.4 - 07/28/08, 4.0.4.5 - 05/04/09. Intends to provide clients with an interface that links all clinical laboratory departments throughout the enterprise.
Recall
- Recall Number
- Z-0365-2011
- Event Number
- 56867
- Firm
- SCC Soft Computer
- FEI Number
- 1000307651
- Product Code
- JQP
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- April 1, 2009
- Posted
- November 16, 2010
- Terminated
- August 7, 2013
- Address
- 5400 Tech Data Drive, Clearwater, FL, 33760
Description
SoftLab GUI Version 4.0.1.0 - 4.0.1.16, 4.0.2.0 - 4.0.2.10, 4.0.3.0 - 4.0.3.13, and 4.0.4.0 - 4.0.4.5. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. Manufacture/Distribution Dates: 4.0.1.0 - 08/06/04, 4.0.1.4 - 10/18/04, 4.0.1.5 - 01/18/04, 4.0.1.6 - 12/15/04, 4.0.1.7 - 01/10/05, 4.0.1.8 - 03/29/05, 4.0.1.9 - 05/06/05, 4.0.1.13 - 01/25/06, 4.0.1.14 - 03/26/07 4.0.1.15 - 05/02/07, 4.0.1.16 - 08/22/08, 4.0.2.0 - 04/08/05, 4.0.2.1 - 08/13/08, 4.0.2.2 - 06/06/05, 4.0.2.3 - 09/19/05, 4.0.2.4 - 09/19/05, 4.0.2.5 - 09/19/05, 4.0.2.7 - 02/13/06, 4.0.2.8 - 12/12/06, 4.0.2.9 - 02/20/08, 04.2.10 - 08/13/08, 4.0.3.2 - 05/17/06, 4.0.3.3 - 09/18/06, 4.0.3.4 - 09/27/06, 4.0.3.5 - 11/29/06, 4.0.3.6 - 06/22/07, 4.0.3.7 - 03/28/07, 4.0.3.8 - 06/22/07, 4.0.3.9 - 08/03/07, 4.0.3.10 - 08/24/07, 4.0.3.11 - 09/24/07, 4.0.3.12 - 03/07/08, 4.0.3.13 - 06/13/08, 4.0.4.2 - 02/22/08, 4.0.4.3 - 05/22/08, 4.0.4.4 - 07/28/08, 4.0.4.5 - 05/04/09. Intends to provide clients with an interface that links all clinical laboratory departments throughout the enterprise.
Clients using Patient Maintenance to perform Moves of Stays may send the wrong billing number during an ADT update to SoftWeb, SoftPath or SoftMedia. The wrong billing number can come from a different patient in Order Entry. Only Moves of Stays are affected.
Clients were notified via the proprietary communication software on 06/01/2009. This is a MANDATORY CORRECTION. The Risk-To-Health Notification Task gives the background on the issue, the potential hazard, root cause, corrective action, alternative working solution, and requested action. Clients were instructed to acknowledge receipt of the notification. If the client wishes to receive the correction, they should document a request to load the software.
Nationwide Distribution
49