FDA Recall Terminated

Advanced Medical Optics (AMO), VISX WaveScan WaveFront System (software). Part Number 0070-1478. Advanced Medical Optics, Santa Clara, CA 95051.

Recall: Z-0365-2008 · Initiated May 30, 2007

Recall

Recall Number
Z-0365-2008
Event Number
39353
Firm
Visx Inc
FEI Number
1221842
Product Code
LZS
Status
Terminated
Root Cause
Software design
Initiated
May 30, 2007
Posted
January 23, 2008
Terminated
April 17, 2008
Address
3400 Central Expressway, Santa Clara, CA, 95051

Description

Advanced Medical Optics (AMO), VISX WaveScan WaveFront System (software). Part Number 0070-1478. Advanced Medical Optics, Santa Clara, CA 95051.

Reason

Erroneous Treatment Calculations - Two software caused errors; in combination, in the WaveScan System Software, will result in an erroneous treatment calculation (overcorrection) in patients. (1) Installation of software for a certain brand of computer (Dedicated Computing LLC) can cause a software registry setting for an algorithm that identifies image reflection to be erroneously set in the "off

Action

AMO sent an Action Required notification on May 30, 2007, to AMO Certified Physicians. The letter included a CD software update (0070-1721) to be installed by users that provided correct registry parameter settings. Upon completion of installation, a confirmation form will print out. Customers are requested to fax back confirmation to AMO (408-773-7055).

Distribution

Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO,MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK,OR,PA, SC, TN, TX, UT, VA, WA, WI, and WV and countries of Canada and Mexico.

Quantity

196 systems are affected in the U.S. and 133 outside of the U.S.