Patient Support Material Number: 9890 010 87431 used with: BuckyDiagnost ; Model: 704060/706037
Recall
- Recall Number
- Z-0364-2016
- Event Number
- 69090
- Firm
- Philips Medical Systems, Inc.
- FEI Number
- 1218950
- Product Code
- KPR
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- August 19, 2014
- Terminated
- June 14, 2016
- Address
- 3000 Minuteman Rd, Andover, MA, 01810-1032
Description
Patient Support Material Number: 9890 010 87431 used with: BuckyDiagnost ; Model: 704060/706037
The hook does not securely hold the footplate in vertical position. Therefore the footplate can fall or swing down and may hurt the operator or patient.
The Field Safety Notice FCO70600086 will be mailed on August 19, 2014. Philips Medical Systems will send out certified letters and will have the receipts returned for tracking purposes as well as have a Field Service Engineer visit each site to provide the fix when the Field Change Order is released.
Distributed in the states of MN, KY, MA, VA, TX, NC, HI, FL, AZ, CT, and WY and the countries of Argentina, Australia, Austria, Canada, Denmark, Germany, Netherlands, Russia, Saudi Arabia, Singapore, Spain, and Switzerland.
1517 systems, not all units have the patient support accessory