Irrigation Tubing for KARL STORZ Endomat Select UP210 REF 031523-10 Tubing Set, Irrigation, PC
Recall
- Recall Number
- Z-0363-2025
- Event Number
- 95607
- Firm
- Karl Storz Endoscopy
- FEI Number
- 3007475226
- Product Code
- NVT
- Status
- Open, Classified
- Root Cause
- Labeling design
- Initiated
- October 21, 2024
- Posted
- November 12, 2024
- Address
- 2151 E Grand Ave, El Segundo, CA, 90245-5017
Description
Irrigation Tubing for KARL STORZ Endomat Select UP210 REF 031523-10 Tubing Set, Irrigation, PC
Due to improper labeling of products. An intended use on the label has not been reviewed and approved by FDA for distribution in the U.S.
On 10/22/2024, an "URGENT: MEDICAL DEVICE CORRECTION" Letters will be sent via email by KARL STORZ informing customers that the firm has become aware that one lot number of misbranded product was distributed to 19 customers. Customers are instructed to: 1. Pass on this Urgent Medical Device Correction to all users of the products listed above (Table 1) and all other persons who need to be aware within your organization. 2. Communicate to all applicable users that the product is not to be utilized during spinal procedures. 3. If you have or may have distributed the products listed, please identify and promptly notify those recipients of this notice. 4. Return the completed Customer Reply Form by E-Mail to the indicated contact below within 15 calendar days from the date of receipt. 5. Please report any incidents related to this issue to the manufacturer, dealer, or local representative and, if applicable, to the national competent authority, as this is important feedback. for questions send an email to [email protected] .
U.S. Nationwide distribution in the states of CO, CT, DC, MD, MI, MN, NC, NY, OH, OK, TN, TX, WA.
40 units