FDA Recall Terminated

SoftPath GUI versions: 4.2.3.3-4.2.3.6 and 4.3.0.5-4.3.0.12. SoftPath is a laboratory information system to be used in a clinical laboratory by knowledgeable, trained, and experienced personnel. SoftPath enables the user to document steps in the receiving and processing of surgical, gynecological, medical cytological, and autopsy case specimens within the Pathology labs, for the purpose of generating Pathology reports for requesting physicians and label creation for identification of specimen/case material.

Recall: Z-0363-2012 · Initiated July 20, 2011

Recall

Recall Number
Z-0363-2012
Event Number
60458
Firm
SCC Soft Computer
FEI Number
1000307651
Product Code
JQP
Status
Terminated
Root Cause
Software design
Initiated
July 20, 2011
Posted
December 2, 2011
Terminated
May 21, 2013
Address
5400 Tech Data Drive, Clearwater, FL, 33760

Description

SoftPath GUI versions: 4.2.3.3-4.2.3.6 and 4.3.0.5-4.3.0.12. SoftPath is a laboratory information system to be used in a clinical laboratory by knowledgeable, trained, and experienced personnel. SoftPath enables the user to document steps in the receiving and processing of surgical, gynecological, medical cytological, and autopsy case specimens within the Pathology labs, for the purpose of generating Pathology reports for requesting physicians and label creation for identification of specimen/case material.

Reason

On 07/20/2011 SCC Soft Computer, Clearwater, FL, initiated a correction on the following SoftPath GUI versions: 4.2.3.3-4.2.3.6 and 4.3.0.5-4.3.0.12. A client reported that the HPV results on a patient report were incorrect. The HPV results were from the previous order on the patient and not the current order. All affected clients were notified about the issue on 07/20/2011.

Action

SCC Soft Computer sent a "CORRECTION COMMUNICATION" dated July 20, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. SCC personnel will review the customized Lab Results AP macro and/or the Micro Results AP macro on customer systems and will either discuss with them the irregularities requiring change or reconfigure macros as necessary.

Distribution

Worldwide Distribution-USA (nationwide) and the country of Canada.

Quantity

97