FDA Recall Terminated

Roche brand Omni S4 Blood Gas Analyzer; Catalog # 03337138001.

Recall: Z-0362-06 · Initiated December 9, 2005

Recall

Recall Number
Z-0362-06
Event Number
34180
Firm
Roche Diagnostics Corp.
FEI Number
1823260
Product Code
CGA
Status
Terminated
Root Cause
Other
Initiated
December 9, 2005
Posted
January 7, 2006
Terminated
April 10, 2006
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

Roche brand Omni S4 Blood Gas Analyzer; Catalog # 03337138001.

Reason

Potential for hemoglobin derivatives which are outside of the quality control (QC) limits to not engage the auto lock out feature of the system for 02Hb, COHb, HHb, MetHb or SO2.

Action

Consignees were notified via letter dated 12/9/05.

Distribution

California, Florida, Georgia, Indiana, Iowa, New Hampshire, Michigan, New York, North Carolina, North Dakota, Pennsylvania, Texas and Washington.