FDA Recall Terminated

Salicylate-SL, in vitro diagnostic, Cat. No.511-20, Lot No. 29892

Recall: Z-0361-2007 · Initiated December 5, 2006

Recall

Recall Number
Z-0361-2007
Event Number
37011
Firm
Diagnostic Chemicals, Ltd.
FEI Number
3002806809
Product Code
DKJ
Status
Terminated
Root Cause
Other
Initiated
December 5, 2006
Posted
January 23, 2007
Terminated
January 27, 2016
Address
16 McCarville Street, Charlottetown Canada Prince Edward Island

Description

Salicylate-SL, in vitro diagnostic, Cat. No.511-20, Lot No. 29892

Reason

Upon subsequent testing on different instruments (as required by other customers), DCL detected a non-linear response on certain analyzers. A decision was made to recall the product due to the fact that DCL customers sell the product to different end users, DCL cannot be certain which instruments the end users employ in their laboratories

Action

For product shipped by DCL to one customer notification letter dated 12/5/06 sent out to customer to return/discard product. Replacement will be sent to customer. For product shipped to DCL (USA) warehouse notification sent to discard kits remaining in stock. For product remaining at DCL Canada in-house notification sent out to discard in-house kits.

Distribution

PA

Quantity

55 kits shipped to one customer