FDA Recall Terminated

Sorting-Drive Professional Part, Sorting-Drive Professional Kit, (Sorting Drive Software v4.2.1), Part No. A94144, Beckman Coulter, Inc. The Sorting-Drive Professional Kit (PN A94144) includes the Sorting-Drive 4.2.1 User Kit (PN B95347) which includes the Sorting-Drive 4.2.1 Software DVD (PN B95346). This software version includes an option for manual barcode printing for sample tubes. Sorting-Drive software facilitates the Laboratory Information System (LIS) connection to AutoMate 2500 Family systems.

Recall: Z-0359-2019 · Initiated May 24, 2018

Recall

Recall Number
Z-0359-2019
Event Number
80434
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
JQP
Status
Terminated
Root Cause
Software change control
Initiated
May 24, 2018
Terminated
December 8, 2022
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

Sorting-Drive Professional Part, Sorting-Drive Professional Kit, (Sorting Drive Software v4.2.1), Part No. A94144, Beckman Coulter, Inc. The Sorting-Drive Professional Kit (PN A94144) includes the Sorting-Drive 4.2.1 User Kit (PN B95347) which includes the Sorting-Drive 4.2.1 Software DVD (PN B95346). This software version includes an option for manual barcode printing for sample tubes. Sorting-Drive software facilitates the Laboratory Information System (LIS) connection to AutoMate 2500 Family systems.

Reason

Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes.Sorting Drive software version 4.2 was replaced by 4.2.1 (both versions contain the same option for manual barcode printing). When printing consecutively numbered tube labels using this optional Barcode Counter function, the system incorrectly converts numbers with consecutive multiple zeroes in the middle, potentially leading to wrong barcode IDs. The firm reports that this does not affect automated label creation or reprinting of individual labels and that no other system functions are affected.

Action

On 6/4/2018 Beckman Coulter (BEC) reported that the issue will be corrected via a field modification performed by Field Service. BEC reported that they do not intend to send a customer letter for this issue.US customers can contact the Technical Support Troubleshooting Hotline at (800) 854-3633.

Distribution

Worldwide Distribution: US (nationwide) and countries of: Algeria, Argentina, Austria, Belgium, Brazil, Chile, China, Czech Republic, Denmark ,France, Germany, Greece ,Hong Kong, India, Ireland, Israel, Italy ,Republic of Korea, Lebanon, Lithuania ,Malaysia, Netherlands, Poland, Portugal , Qatar, Romania, Russian Federation, Saudi Arabia ,South Africa , Spain, Sweden, Taiwan, Turkey , and United Kingdom.

Quantity

58