Power supply boards on Dash 3000, 4000 and 5000 Monitors. GE Healthcare, 8200 West Tower Avenue, Milwaukee, WI 53223.

Recall: Z-0359-2011 · Initiated October 11, 2010

Recall

Recall Number
Z-0359-2011
Event Number
56664
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
MHX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
October 11, 2010
Posted
November 15, 2010
Terminated
May 29, 2012
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Power supply boards on Dash 3000, 4000 and 5000 Monitors. GE Healthcare, 8200 West Tower Avenue, Milwaukee, WI 53223.

Reason

The power supply board in GE Healthcare Dash 3000, 4000 and 5000 patient monitors may fail due to insufficient solder of a component. Insufficient solder on the power supply board could result in excess EMC emissions, smoking of components and eventual failure of the power supply, tripping of the fuse and immediate shutdown resulting in a loss of monitoring that could delay recognition of a clinic

Action

GE Healthcare issued an "Urgent Medical Device Correction" letter dated October 8, 2010, which described the loss of monitoring issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. Customers were instructed to remove the units from use. Local service representatives will contact firms to schedule installation of replacement power supply boards. GE Healthcare can be contacted at 1-800-558-7044, Option 2, Technical Support, about this correction.

Distribution

Worldwide Distribution: USA in the states of IL, NV, TX, and VA, and the countries of Australia, Brazil, China, Japan, South Korea, and Taiwan.

Quantity

137