FDA Recall Open, Classified

GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Part Numbers 00024307504, 00024330004, 00024345004

Recall: Z-0357-2024 · Initiated September 29, 2023

Recall

Recall Number
Z-0357-2024
Event Number
93313
Firm
Instrumentation Laboratory
FEI Number
1217183
Product Code
CHL
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
September 29, 2023
Posted
November 21, 2023
Address
180 Hartwell Rd, Bedford, MA, 01730-2443

Description

GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Part Numbers 00024307504, 00024330004, 00024345004

Reason

Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.

Action

An URGENT MEDICAL DEVICE RECALL notification letter dated 9/29/23 was sent to customers. Mandatory Customer Actions Please take the following immediate actions: " Check your inventory for the GEM Premier 3000 Non-iQM PAK (cartridge) lot numbers listed in Appendix A. " Discontinue use and destroy any GEM Premier 3000 Non-iQM PAK (cartridge) lot numbers listed in Appendix A. " Document the destruction on the Mandatory Tracking Response Record and return the completed and signed form to Werfen Regulatory at the fax number or e-mail address listed below. " Contact Werfen Customer Service at 1-800-955-9525 to order replacement PAKs (cartridges). " Share this information with your laboratory staff and follow your internal procedures. " Forward this notification to all affected locations within your facility. " Retain a copy of this notification for your records. Mandatory Response Tracking Record " Return the completed and signed Mandatory Response Tracking Form to: " Fax No.: 1-781-861-4207 Or " E-mail: [email protected] Contact Information for Questions " For technical questions, contact the Technical Support Center at 1-800-678-0710. " For replacement cartridges, contact Customer Service at 1-800-955-9525.

Distribution

US States: MA, MN, NY. India.

Quantity

2448 total