FDA Recall Terminated

Boston Scientific brand Neuro Renegade Hi-Flo Microcatheter, 150cm/20cm; Rx Only; Model Numbers: 182880US, 182890US, 182900US, 182910US; Product is distributed by Boston Scientific Corporation, Natick, MA; Product is manufactured by Boston Scientific Corporation, Cork, Ireland

Recall: Z-0357-2011 · Initiated August 24, 2009

Recall

Recall Number
Z-0357-2011
Event Number
53048
Firm
Boston Scientific Corporation
FEI Number
2939204
Product Code
KRA
Status
Terminated
Root Cause
Packaging process control
Initiated
August 24, 2009
Posted
November 15, 2010
Terminated
February 3, 2011
Address
47900 Bayside Pkwy, Fremont, CA, 94538-6515

Description

Boston Scientific brand Neuro Renegade Hi-Flo Microcatheter, 150cm/20cm; Rx Only; Model Numbers: 182880US, 182890US, 182900US, 182910US; Product is distributed by Boston Scientific Corporation, Natick, MA; Product is manufactured by Boston Scientific Corporation, Cork, Ireland

Reason

Sterility Loss-- The sterile barrier pouch seal (top seal) may be breached by the product, resulting in loss of sterility which may result in transmission of disease or infection.

Action

Boston Scientific issued an Urgent Medical Device Recall letter dated August 26, 2009 identifying the affected devices, the issue prompting the recall, and actions to be taken by customers. Customers were instructed to determine if they have affected product in their inventory, and if so, immediately quarantine and return to Boston Scientific. Customers can contact Boston Scientific at 510 624-1734.

Distribution

Distribution Worldwide, including throughout the US.