FDA Recall Terminated

Contact Detach, 8mm, 23", G29; Contact Detach 8 mm, 43", G29; Contact Detach, 6mm, 23" G29; Infusion Set, Contact Detach, 6mm x 23"; Infusion Set, Contact Detach, 6mm x 32"; Infusion Set, Contact Detach, 8mm x 23"; Infusion Set, Contact Detach, 8mm x 32" The products are used in combination with insulin infusion pumps for continuous subcutaneous infusion of insulin

Recall: Z-0356-2017 · Initiated November 7, 2014

Recall

Recall Number
Z-0356-2017
Event Number
74242
FEI Number
1000477110
Product Code
FPA
Status
Terminated
Root Cause
Device Design
Initiated
November 7, 2014
Terminated
November 16, 2016
Address
Unomedical As Aholmvej 1 3 Osted Roskilde Denmark

Description

Contact Detach, 8mm, 23", G29; Contact Detach 8 mm, 43", G29; Contact Detach, 6mm, 23" G29; Infusion Set, Contact Detach, 6mm x 23"; Infusion Set, Contact Detach, 6mm x 32"; Infusion Set, Contact Detach, 8mm x 23"; Infusion Set, Contact Detach, 8mm x 32" The products are used in combination with insulin infusion pumps for continuous subcutaneous infusion of insulin

Reason

Tubing detachment

Action

ConvaTec sent an Urgent Medical Device Correction letter to all affected consignees on November 7, 2014. The letter notified the consignees of what action they need to take if they experience a detached tubing. You can continue to use your infusion sets by following the advice below. 1. When changing your infusion set, closely follow the instructions for use included in the product box. Check the tubing at the connect/disconnect location identified in the drawings above to make sure it is not loose. 2. As always, it is essential to monitor your blood sugar levels frequently using your blood glucose meter. Proactively check your tubing connections occasionally throughout the day to ensure tubing is secure. It is especially important to check your blood sugar and tubing connections at bedtime to confirm insulin delivery is occurring. 3. If you experience a high blood sugar, check your tubing connections and infusion site closely to ensure your tubing is secure. If you discover the tubing is detached: 1. Do not attempt to reattach the tubing. Replace the infusion set immediately. 2. Treat any high blood sugar based on guidelines provided by your healthcare professional. 3. Call your distributor's Helpline to report the issue. They will give you instructions on how to return the affected infusion set. The letter included a reply card to be returned.

Distribution

Worldwide Distribution - US including CA, IN, PA, and Internationally to Canada, Korea, and Japan.

Quantity

664,000 units