FDA Recall Terminated

ARTAS Disposable and Reuseable Kit ( Contains Skin Tensioner FP-45384; Inner Needle with Vacuum Tube FP-45385; dissection Punch 4.0mm FP 23304; Saline Nose Cone FP-45387; Follicle Trap FP 23306; Non-Sterile Disposable kit FP 23267 and Molded Parts, Reusable Kit FP 45538 Rx): Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.

Recall: Z-0356-2014 · Initiated October 13, 2013

Recall

Recall Number
Z-0356-2014
Event Number
66558
Firm
Restoration Robotics
FEI Number
3006515340
Product Code
ONA
Status
Terminated
Root Cause
Process control
Initiated
October 13, 2013
Posted
November 18, 2013
Terminated
December 10, 2013
Address
128 Bayview Drive, San Jose, CA, 95134

Description

ARTAS Disposable and Reuseable Kit ( Contains Skin Tensioner FP-45384; Inner Needle with Vacuum Tube FP-45385; dissection Punch 4.0mm FP 23304; Saline Nose Cone FP-45387; Follicle Trap FP 23306; Non-Sterile Disposable kit FP 23267 and Molded Parts, Reusable Kit FP 45538 Rx): Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.

Reason

Reports of discoloration (residue) on the pins of the skin tensioner have been received by several customers. Discoloration due to insufficient cleaning of stainless steel pins during passivation process. The residue has been identified as a compound of iron and chlorine.

Action

Restoration Robotics sent an Urgent Recall letters dated October 13, 2013 to all distributors via first class mail. Direct accounts were notified by phone as well as fax. Customers were informed of the affected product, problem and actions to be taken. Replacement products are being distributed to each site and will be delivered to each customer. For questions call 408-883-6760

Distribution

Worldwide Distribution - USA Nationwide in the countries of: Korea, Japan, Singapore, Thailand, United Kingdom, Canada, and Taiwan.

Quantity

515 US, 315 International (all Packaging configurations.