Infusomat Administration Sets as listed below: 362031 Infusomat Space Pump Set with Caresite Injection Site; 362032 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; 362033 UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; 362034 Priming volume: 19mL, length: 120in, SafeLine; 362035 INFUSOMAT SPACE PUMP IV SET; 362043 Priming volume: 39mL, length: 120in, SafeLine; 362050 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; 362420 SPACE PUMP SET 15 DROP W/2 SAFEDAY; 362432 15 drops/mL, priming volume: 21mL, length: 118in, ULTRASITE; 362911 Priming volume: 14mL, length: 84in, CARESITE, Spin-Lock; 363010 Priming volume: 21mL, length: 121in, SafeLine; 363019 Priming volume: 40mL, length: 123in, SafeLine; 363030 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; 363032 20 drops/mL, priming volume: 20mL, length: 119in, SafeLine; 363230 15 drops/mL, priming volume: 22mL, length: 123in, ULTRASITE; 363410 15 drops/mL, priming volume: 23mL, length: 119in, CARESITE; 363419 Priming volume: 41mL, length: 120in, CARESITE; 363420 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; 363421 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; 363422 Priming volume: 21mL, length: 130in, CARESITE; 363423 60 drops/mL, priming volume: 21mL, length: 118in, CARESITE; 363424 UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; 363430 15 drops/mL, priming volume: 23mL, length: 123in, CARESITE; 363433 UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; 363901 60 drops/mL, priming volume: 11mL, length: 129in; 363902 Priming volume: 25mL, length: 130in; 363904 UNIV. 15 DROP PUMP SET, ASV; 470119 15 drops/mL, priming volume: 23mL, length: 119in, CARESITE; 480254 SPACE PUMP IV SET 15D, 110 IN.; 480255 SPACE PUMP IV SET, 127 IN; 480257 60 DROP NITRO SPACE SET, 88 IN.; 480258 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; 480263 SPACE PUMP IV SET 15D, 90 IN.; 480264 SPACE PUMP IV SET, 15 DRPS/ML, 95 IN.; 490037 Priming volume: 25mL, length: 136in; 490038 60 drops/mL, priming volume: 11mL, length: 129in; 490100 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; 490101 Priming volume: 19mL, length: 120in, CARESITE; 490102 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; 490103 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; 490104 60 drops/mL, priming volume: 21mL, length: 115in, CARESITE; 490105 Priming volume: 39mL, length: 120in, CARESITE
Recall
- Recall Number
- Z-0355-2024
- Event Number
- 93232
- Firm
- B. Braun Medical, Inc.
- FEI Number
- 2523676
- Product Code
- FPA
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- November 1, 2023
- Posted
- December 1, 2023
- Address
- 901 Marcon Blvd, Allentown, PA, 18109-9512
Description
Infusomat Administration Sets as listed below: 362031 Infusomat Space Pump Set with Caresite Injection Site; 362032 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; 362033 UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; 362034 Priming volume: 19mL, length: 120in, SafeLine; 362035 INFUSOMAT SPACE PUMP IV SET; 362043 Priming volume: 39mL, length: 120in, SafeLine; 362050 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; 362420 SPACE PUMP SET 15 DROP W/2 SAFEDAY; 362432 15 drops/mL, priming volume: 21mL, length: 118in, ULTRASITE; 362911 Priming volume: 14mL, length: 84in, CARESITE, Spin-Lock; 363010 Priming volume: 21mL, length: 121in, SafeLine; 363019 Priming volume: 40mL, length: 123in, SafeLine; 363030 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; 363032 20 drops/mL, priming volume: 20mL, length: 119in, SafeLine; 363230 15 drops/mL, priming volume: 22mL, length: 123in, ULTRASITE; 363410 15 drops/mL, priming volume: 23mL, length: 119in, CARESITE; 363419 Priming volume: 41mL, length: 120in, CARESITE; 363420 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; 363421 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; 363422 Priming volume: 21mL, length: 130in, CARESITE; 363423 60 drops/mL, priming volume: 21mL, length: 118in, CARESITE; 363424 UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; 363430 15 drops/mL, priming volume: 23mL, length: 123in, CARESITE; 363433 UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; 363901 60 drops/mL, priming volume: 11mL, length: 129in; 363902 Priming volume: 25mL, length: 130in; 363904 UNIV. 15 DROP PUMP SET, ASV; 470119 15 drops/mL, priming volume: 23mL, length: 119in, CARESITE; 480254 SPACE PUMP IV SET 15D, 110 IN.; 480255 SPACE PUMP IV SET, 127 IN; 480257 60 DROP NITRO SPACE SET, 88 IN.; 480258 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; 480263 SPACE PUMP IV SET 15D, 90 IN.; 480264 SPACE PUMP IV SET, 15 DRPS/ML, 95 IN.; 490037 Priming volume: 25mL, length: 136in; 490038 60 drops/mL, priming volume: 11mL, length: 129in; 490100 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; 490101 Priming volume: 19mL, length: 120in, CARESITE; 490102 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; 490103 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; 490104 60 drops/mL, priming volume: 21mL, length: 115in, CARESITE; 490105 Priming volume: 39mL, length: 120in, CARESITE
In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.
On November 1, 2023, the firm notified customers via URGENT MEDICAL DEVICE CORRECTION letters. Customers were informed that administration sets with lot number 0061902124 and higher were manufactured with revised specifications which should reduce the potential for false air-in-line alarms. After November 30, 2023, all lots distributed by BBMI will have been manufactured with these revised specifications. The firm provided interim measures for users to apply for affected product. For example, users should not use tubing with noticeably smaller outer diameter or which needs to be stretched to fit within the pump. If possible, unaffected lots of product should be diverted to areas that infuse high risk medications or other infusions with acute clinical effects due to initial limited supply of the new administration sets. Affected lots should be diverted to care areas that do not infuse high risk medications or other infusions with acute clinical effects. When infusing high risk medications or infusions with acute clinical effects, customers were advised to have a second pump with an administration set available for exchange in the event of a false alarm. If a second pump is not available, have a second administration set available for exchange in the event of a a false alarm. Additional patient monitoring is required. Adverse reactions or quality problems experienced with this product, or questions about this correction may be reported to BBMI s Postmarket Surveillance Department by calling 1-833-425-1464.
Distribution US nationwide, Canada and Germany.
33,820,976 units