ARTAS Disposable Kit ( Contains Skin Tensioner FP-45384; Inner Needle with Vacuum Tube FP-45385; Dissection Punch 4.0mm FP-23304; Saline Nose Cone FP-45387; Follicle Trap - FP-23306 and Non-sterile Disposable Clinical Kit FP -23267): and ARTAS Disposable Kits, 18 Gauge Epsilon (Contains Disposable Kit, 18 Gauge Epsilon FP 45547 and Non-Sterile Disposable Kit FP-23267). RX. Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.
Recall
- Recall Number
- Z-0354-2014
- Event Number
- 66558
- Firm
- Restoration Robotics
- FEI Number
- 3006515340
- Product Code
- ONA
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- October 13, 2013
- Posted
- November 18, 2013
- Terminated
- December 10, 2013
- Address
- 128 Bayview Drive, San Jose, CA, 95134
Description
ARTAS Disposable Kit ( Contains Skin Tensioner FP-45384; Inner Needle with Vacuum Tube FP-45385; Dissection Punch 4.0mm FP-23304; Saline Nose Cone FP-45387; Follicle Trap - FP-23306 and Non-sterile Disposable Clinical Kit FP -23267): and ARTAS Disposable Kits, 18 Gauge Epsilon (Contains Disposable Kit, 18 Gauge Epsilon FP 45547 and Non-Sterile Disposable Kit FP-23267). RX. Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.
Reports of discoloration (residue) on the pins of the skin tensioner have been received by several customers. Discoloration due to insufficient cleaning of stainless steel pins during passivation process. The residue has been identified as a compound of iron and chlorine.
Restoration Robotics sent an Urgent Recall letters dated October 13, 2013 to all distributors via first class mail. Direct accounts were notified by phone as well as fax. Customers were informed of the affected product, problem and actions to be taken. Replacement products are being distributed to each site and will be delivered to each customer. For questions call 408-883-6760
Worldwide Distribution - USA Nationwide in the countries of: Korea, Japan, Singapore, Thailand, United Kingdom, Canada, and Taiwan.
515 US, 315 International (all packaging configurations)