2000 Vascular with 30 Degree option
Recall
- Recall Number
- Z-0353-2020
- Event Number
- 83219
- Firm
- Heritage Medical Products, Inc.
- FEI Number
- 3003881889
- Product Code
- LGX
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- September 2, 2016
- Terminated
- December 14, 2020
- Address
- 10380 County Road 6310, West Plains, MO, 65775-6331
Description
2000 Vascular with 30 Degree option
Potential for bed to drop suddenly at the head while a patient is laying on it.
Heritage Medical Products, Inc. notified customers on about 09/02/2016 via "Advisory Notice." Customers were informed that Heritage Medical has designed a reinforcement kit as a correction, which will be supplied to customers. Customers were instructed to contact Heritage Medical, provide the serial number of their bed to ensure it is affected, and Heritage Medical will provide a free kit that will reinforce the area on the affected devices that has experienced failures. The contact information is a follows: Heritage Medical Products, Inc., ATTN.: Trunnion Reinforcement Kit, 10380 CR 6310, West Plains, MO 65775, (888) 256-8748.
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, OH, OK, OR, PA, SC, TX, UT, VA, WI, WY, PR. International distribution to Canada, Saudi Arabia, Trinidad & Tobago.
167 beds (159 US, 8 OUS)