FDA Recall Terminated

Epsilon Skin Tensioner:Rx, Sterile Sterilized using Gamma. Turns Red if sterilized using Gamma; Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.

Recall: Z-0353-2014 · Initiated October 13, 2013

Recall

Recall Number
Z-0353-2014
Event Number
66558
Firm
Restoration Robotics
FEI Number
3006515340
Product Code
ONA
Status
Terminated
Root Cause
Process control
Initiated
October 13, 2013
Posted
November 18, 2013
Terminated
December 10, 2013
Address
128 Bayview Drive, San Jose, CA, 95134

Description

Epsilon Skin Tensioner:Rx, Sterile Sterilized using Gamma. Turns Red if sterilized using Gamma; Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.

Reason

Reports of discoloration (residue) on the pins of the skin tensioner have been received by several customers. Discoloration due to insufficient cleaning of stainless steel pins during passivation process. The residue has been identified as a compound of iron and chlorine.

Action

Restoration Robotics sent an Urgent Recall letters dated October 13, 2013 to all distributors via first class mail. Direct accounts were notified by phone as well as fax. Customers were informed of the affected product, problem and actions to be taken. Replacement products are being distributed to each site and will be delivered to each customer. For questions call 408-883-6760

Distribution

Worldwide Distribution - USA Nationwide in the countries of: Korea, Japan, Singapore, Thailand, United Kingdom, Canada, and Taiwan.

Quantity

515 in US. 315 International.(included all packaging configurations)