Baxter V-Link Luer Activated Device with Vitalshield Protective Coating Non-DEHP Catheter Extension Set Kit; 7.6" (19.3 cm), Vol. 1.0 mL, 1 Catheter Extension Set with Luer Activated Device with Silver for IV Access and Male Luer Lock Adapter with Retractable Collar, 1 VITAL-HOLD medium catheter stabilization device, 2 foam tape strips; an Rx sterile convenience kit for IV infusions; 50 units per case; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60015 U.S.A., Made in Costa Rica; catalog 6N8220
Recall
- Recall Number
- Z-0351-2011
- Event Number
- 56792
- Firm
- Baxter Healthcare Corp. Rt.
- FEI Number
- 1416980
- Product Code
- LBJ
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- September 17, 2010
- Posted
- November 15, 2010
- Terminated
- October 4, 2011
- Address
- 120 & Wilson Rd, Round Lake, IL, 60073
Description
Baxter V-Link Luer Activated Device with Vitalshield Protective Coating Non-DEHP Catheter Extension Set Kit; 7.6" (19.3 cm), Vol. 1.0 mL, 1 Catheter Extension Set with Luer Activated Device with Silver for IV Access and Male Luer Lock Adapter with Retractable Collar, 1 VITAL-HOLD medium catheter stabilization device, 2 foam tape strips; an Rx sterile convenience kit for IV infusions; 50 units per case; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60015 U.S.A., Made in Costa Rica; catalog 6N8220
The removable paper liner covering the adhesive strip that is used to secure the catheter stabilization device to the patient is pyrogenic. The device is labeled as nonpyrogenic.
Baxter issued Urgent Product Recall letters dated September 17, 2010 to the affected direct accounts. The accounts were informed that the paper lining covering the adhesive strip that is used to secure the catheter stabilization device to the patient was found to be pyrogenic. The accounts were requested to discontinue use and segregate the affected lots of products from inventory and call Baxter Healthcare Center for Service at 1-888-229-0001, Monday through Friday, 7:00 AM to 6:00 PM CT, to arrange for the return of the affected product for credit. The accounts were also requested to complete and fax back to Baxter the enclosed reply sheet confirming receipt of the letter and the amount of each product being returned. If the accounts further distributed the products, they were requested to forward the communication to those accounts. Clinical questions regarding the communication are directed to the Medical Information Services at Baxter at 1-800-933-0303. Any other questions may be directed to the Center for One Baxter at 1-800-422-9837.
Nationwide Distribution: Throughout the US.
73,350 units