FDA Recall Terminated

Jostra Heater Cooler Unit HCU30, system version 1.02-Low voltage machine

Recall: Z-0350-2007 · Initiated November 10, 2006

Recall

Recall Number
Z-0350-2007
Event Number
37023
Firm
MAQUET Inc.
FEI Number
3008355164
Product Code
DWC
Status
Terminated
Root Cause
Other
Initiated
November 10, 2006
Posted
January 25, 2007
Terminated
February 5, 2007
Address
1140 US Highway 22, Bridgewater, NJ, 08807-2958

Description

Jostra Heater Cooler Unit HCU30, system version 1.02-Low voltage machine

Reason

The HCU 30 system potentially will not be have the power capacity to run the compressor when the heaters are on during operation. This may lead to a situation where the user may not be able to cool the patient and/or cardioplegia sufficiently.

Action

Letters were mailed to customers on 11/10/2006 by Federal Express.

Distribution

Hospitals in CA, NC, FL, WY, NY and WV.S

Quantity

13 units