FDA Recall Terminated

Model 055: Thermophore Classic, Professional Quality, Deep-Heat Therapy, Model 055 Large 14' x 27", Model 056: Thermophore Classic, Professional Quality Deep-Heat Therapy, Model 056 Medium 14" x 14", Model 077: Thermophore Classic Professional Quality Deep-Heat Therapy, Model 077 Petite 4"x 17",Model 095: Thermophore Classic, Professional Quality Deep-Heat Therapy, Model 095 Large 14" x 27 Model 096: Thermophore Classic, Professional Quality Deep-Heat therapy Model 096, Medium, 14" x 14" Model 097: Thermophore Classic, Professional quality Deep-Heat Therapy, Model 097 , Petite, 4" x 17", Model 255: Thermophore Classic Plus, Professional Quality Deep-Heat Therapy, Model 255 Large,14" x 27", Model 256: Thermophore Classic Plus Professional Quality Deep-Heat Therapy, Model 256, Medium, 14" X 14 Model 277: Thermophore Classic Plus Professional Quality Deep-Heat Therapy, Model 277, Petite , 4" x 17" Used for the temporary relief of joint and muscle pain.

Recall: Z-0349-2017 · Initiated September 15, 2016

Recall

Recall Number
Z-0349-2017
Event Number
73286
Firm
Battle Creek Equipment Co.
FEI Number
1833125
Product Code
IRT
Status
Terminated
Root Cause
Device Design
Initiated
September 15, 2016
Terminated
January 11, 2018
Address
702 S Reed Rd, Fremont, IN, 46737-2098

Description

Model 055: Thermophore Classic, Professional Quality, Deep-Heat Therapy, Model 055 Large 14' x 27", Model 056: Thermophore Classic, Professional Quality Deep-Heat Therapy, Model 056 Medium 14" x 14", Model 077: Thermophore Classic Professional Quality Deep-Heat Therapy, Model 077 Petite 4"x 17",Model 095: Thermophore Classic, Professional Quality Deep-Heat Therapy, Model 095 Large 14" x 27 Model 096: Thermophore Classic, Professional Quality Deep-Heat therapy Model 096, Medium, 14" x 14" Model 097: Thermophore Classic, Professional quality Deep-Heat Therapy, Model 097 , Petite, 4" x 17", Model 255: Thermophore Classic Plus, Professional Quality Deep-Heat Therapy, Model 255 Large,14" x 27", Model 256: Thermophore Classic Plus Professional Quality Deep-Heat Therapy, Model 256, Medium, 14" X 14 Model 277: Thermophore Classic Plus Professional Quality Deep-Heat Therapy, Model 277, Petite , 4" x 17" Used for the temporary relief of joint and muscle pain.

Reason

Products manufactured with the SPT cord have a known failure mode of sparking,causing smoke, fire and burns.

Action

All consumers for which Battle Creek has contact information and that purchased a recalled pad will be contacted via first class mail. Consumers will be asked to contact us by phone after verifying that the model number and lot code is part of the recall. Battle Creek plans to add information about the recall on its web site including a pdf of the consumer letter. All customer consignees (non-consumer) will be informed of the recall. Email will be the most effective method , and mail and telephone calls will be used as needed.

Distribution

Distributed within the United States, Canada, Mexico, Puerto Rico and Columbia

Quantity

99,060