FDA Recall Terminated

Platelia Aspergillus EIA

Recall: Z-0348-04 · Initiated November 6, 2003

Recall

Recall Number
Z-0348-04
Event Number
28013
Firm
Bio Rad Laboratories, Inc.
FEI Number
3022521
Product Code
NOM
Status
Terminated
Root Cause
Other
Initiated
November 6, 2003
Posted
January 14, 2004
Terminated
January 27, 2004
Address
6565 185th Ave Ne, Redmond, WA, 98052-5039

Description

Platelia Aspergillus EIA

Reason

Potential drug and device interaction with piperacillin/tazobactam (Zosyn) causing positive test result for galactomannan.

Action

Customers were sent a letter, Technical Bulletin and revised package insert on November 6, 2003.

Distribution

The firm distributed to hospitals and diagnostic laboratories nationwide.

Quantity

147 kits