Accelerate PhenoTest BC kit REF 10101018
Recall
- Recall Number
- Z-0345-2023
- Event Number
- 91075
- Firm
- Accelerate Diagnostics Inc
- FEI Number
- 3010671651
- Product Code
- PRH
- Status
- Open, Classified
- Root Cause
- Mixed-up of materials/components
- Initiated
- October 17, 2022
- Posted
- December 6, 2022
- Address
- 3950 S Country Club Rd, Ste 470, Tucson, AZ, 85714-2240
Description
Accelerate PhenoTest BC kit REF 10101018
Due improperly formulated and released bulk lot of fluorescence in situ hybridization (FISH) probes resulting in false positive identification results for Staphylococcus aureus and Streptococcus spp.
On 11/04/2022, the firm communicated an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that the firm has become aware of complaints concerning false positive identification results for Staphylococcus aureus and Streptococcus spp. when using the Accerlerate PhenoTest BC kits. Customer are instructed to: (1) Share this customer notification with all who need to be made aware within their organization. (2) Appropriately destroy all affected Accelerate PhenoTest BC kits by defacing all labeling and disposing in accordance with site biohazard or waste policies and procedure. (3) Complete the Recall Response Form and return it to Accelerate Diagnostics. (4) Retrospectively review runs from affected lots, specifically those runs with positive identification results for either Staphylococcus aureus or Streptococcus spp. Follow their institutional policies and processes to review prior patient results related to the described issue. (5) Continue to report all technical issues through Accelerate Technical Support. Contact for Questions: NA Customer Service & Technical Support Toll free: +1 (888) 586 2939 opt. 4 Direct: +1 (520) 365 3100 Email: [email protected]
U.S.: AR, CA, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, MO, MS, MT, NC, ND, NJ, NY, OH, PA, RI, SC, TN, TX, VA, WA, and WV. O.U.S.: Germany, Italy, Kuwait, Latvia, Poland, Portugal, Romania, Saudi Arabia, Spain, and United Arab Emirates
200 kits